Support
FAQ
Common questions answered to help you troubleshoot your Audioscan product.
- Within the Audioscan Noah Module, you will want to open “Network troubleshooting” from the Settings icon located in the bottom left corner of the module.
- Click Settings > Network Troubleshooting > Run All
- If you have entered your IP address correctly within the Audioscan Noah Module, and there is an “X” beside the Ping test, you will want to “Export Results” and provide this information to your IT administrator.
There are two main methods to update the software on your Verifit2 or Axiom
1. Via the Audioscan Noah Module
- Click Settings > Check now for updates
For more information, watch the Audioscan Noah Module screen tour
2. Via USB stick
- Visit audioscan.com for software downloads and software update instructions
For more information, watch the Updating Audioscan System Software video
The simplest method to update your Audioscan Noah Module, is to open the Audioscan Noah Module, click Settings > Check now for updates
For more information, watch the Audioscan Noah Module screen tour
The Audioscan Noah Module can also be downloaded from the Audioscan website. Visit www.audioscan.com for more information
Check Test Box cable and microphones are plugged in properly.
Contact your local distributor for assistance.
You will be prompted to calibrate your probe microphones whenever your calibration has expired. You can manually select the calibration by clicking Main Menu > Calibration within the “On-ear” column of the main menu.
You can follow the instructions on-screen.
To learn more about the probe microphone calibration process, watch the Audioscan Probe Microphone Calibration video.
If you have purchased the Audioscan Noah Module and need your license key, please contact us at: 1-(800)-265-2093 or support@audioscan.com. Please include your Audioscan model and serial number.
The MaxTM SPL (maximum sound pressure level at the tympanic membrane) exceeded prompt is a safety limit to ensure output levels of the hearing aid does not exceed an unsafe level. This safety feature can be adjusted or disabled by clicking Settings > Max TM SPL and selecting the desired MaxTM SPL.
Calibration failure should indicate a faulty microphone or set up. User should swap mics, Left-for-Right, and try to calibrate both sides again. If the fault follows the mic, replacement is required. If neither side will calibrate with either mic, suspect faulty jacks/cabling.
There are a couple reasons why targets might be missing from Speechmap after you have entered or imported audiometric thresholds. You may want to check the following:
- Confirm a calibrated stimulus is selected
- If using NAL-NL2 fitting formula, ensure 500 Hz and 2 kHz thresholds are entered
In some cases, after the Verifit2 has been updated, you may be required to re-establish the network connection. To troubleshoot and resolve the issue, you will want to check the following:
- Confirm your Verifit2 and PC are connected to the same network
- Confirm the correct network address has been entered into the Audioscan Noah Module
- Use the Audioscan Noah Module’s Network Troubleshooting tool to confirm the network has been configured correctly
Contact Us
Please leave your information below and we will get back to you as soon as we can.
Information
Audioscan, A Division of Etymonic Design Inc
20 Ludwig Street, Dorchester, Ontario, N0L 1G4
Tel: 519-268-3313
Tel: 800-265-2093
Monday – Friday 8:00am – 4:30pm EST